Original ArticleInternational Journal of Pharmaceutical InvestigationVol. 13 | Issue 1 | pp. 113–121Open access
Related Substances Method Development and Validation of Axitinib, Zanubrutinib and Upadacitinib Using RP-HPLC and its Degradation Products were Characterized by Using LC-MS/MS
- 1,
- 1*
- 1 Department of Chemistry, JKC College Guntur, Acharya Nagarjuna University, Guntur Andhra Pradesh, INDIA.
Published in International Journal of Pharmaceutical Investigation
Correspondence: Madhavi Nannapaneni
Department of Chemistry, JKC College Guntur, Acharya Nagarjuna University, Guntur Andhra Pradesh, INDIA.
Email: madhavijkcchempg@gmail.com
- Received:
- Jul 23, 2022
- Accepted:
- Oct 12, 2022
- DOI:
- 10.5530/223097131696
How to cite
Syed, I. B., & Nannapaneni, M. Related Substances Method Development and Validation of Axitinib, Zanubrutinib and Upadacitinib Using RP-HPLC and its Degradation Products were Characterized by Using LC-MS/MS. International Journal of Pharmaceutical Investigation, 13(1), 113–121. https://doi.org/10.5530/223097131696
Abstract
Background: Axitinib was used to treat carcinoma of renal cells and Upadacitinib was used to treat rheumatoid arthritis and Zanubrutinib used as an anti cancer medication for mantle cell lymphoma. In the current application, Zanubrutinib, Upadacitinib and Axitinib and their associated substances will be developed and validated. Materials and Methods: The optimised method includes the Zanubrutinib, Upadacitinib and Axitinib gradient elution and associated flow rates with dimensions of 1 mL/min and phenyl column X-bridge (150 x 4.6 mm, 3.5 μ). A mobile phase was employed with 1.2 g of Hexane sulphonic acid of pH-2.5 adjusted with ortho-phosphoric acid (buffer) and acetonitrile. Zanubrutinib, Upadacitinib and Axitinib were separated from their associated substances for a total run time of 60 min. The approach developed has been validated in accordance with the ICH guidelines. Results: The test concentration LOD and LOQ were established for both drugs and for their impurities. The diagrams that were drawn were straightforward with an R2 regression coefficient > 0.999. As part of the method validation, recovery, specificity, linearity, accuracy, robustness was determined and the results were found to be within an acceptable range. Conclusion: HPLC was used to validate the method in terms of accuracy, linearity, method precision, accuracy, limit of detection, limit of quantification, robustness, and degradation, and LC-MS/MS was used to characterise the degradation products. The system was validated in all of these aspects.
Keywords
Subject
Article metadata
| Title | Related Substances Method Development and Validation of Axitinib, Zanubrutinib and Upadacitinib Using RP-HPLC and its Degradation Products were Characterized by Using LC-MS/MS |
|---|---|
| Authors | Ibrahim Baje Syed; Madhavi Nannapaneni |
| Affiliations | Department of Chemistry, JKC College Guntur, Acharya Nagarjuna University, Guntur Andhra Pradesh, INDIA. |
| Corresponding author | madhavijkcchempg@gmail.com |
| Journal | International Journal of Pharmaceutical Investigation |
| Volume / Issue | Vol. 13, Issue 1 |
Also in this issue
- A Review on Different Solubility Enhancement Techniques of Ticagrelorpp. 1–6
- Herbal Bioenhancers with Nanocarriers: A Promising Approach for Oral Peptide Deliverypp. 7–13
- Clozapine: A Scientific Analysis of Global Publications during 1970-2021pp. 14–21
- Cardioprotective and Hepatoprotective Effects of Ixora pavetta and Tecoma capensis Leaf Extractspp. 22–27
- Optimization of Green Synthesized Black Tea Nanoparticles using Central Composite Designpp. 28–36